Skip to main content
sc_it
Visitor II
September 25, 2014
Question

CW half-bridge

  • September 25, 2014
  • 1 reply
  • 723 views
Posted on September 25, 2014 at 09:53

Why did you design a specific CW half-bridge with current limited in 0.6A output current capability?

#ultrasound
This topic has been closed for replies.

1 reply

ricciardo.antonio
Visitor II
April 23, 2015
Posted on April 23, 2015 at 10:42

Hi, Lee,

The CW HB was implemented to keep low the power consumption during the CW operations [CW, PW Doppler and Color Flow]. The fact that we made up customizing the P/N Ch power FETs size [CW HB] allowed us to reduce the parasitic capacitances, hence to save powers during HV commutations.

Regards

Antonio [

mailto:antonio.ricciardo@st.com