Posted on February 09, 2018 at 15:20

Hi all,

I have been working on H3LIS331DL sensor. I am missing one clarification from datasheet of H3LIS331DL.

On Page 10 of datasheet H3LIS331DL, Turn on time should be 1/ODR + 1ms.

It is clear that based on ODR rate we need to set turn on time.

Now my question is this turn on time either we need to provide after power on of the accelerometer or after configuring Control Registers of the accelerometer.

Anyone have any idea please clarify on this.

Thanks,

Siva.

#h3lis331dl-odr #h3lis331dl-configuration #h3lis331
    This topic has been closed for replies.

    1 reply

    ST Employee
    February 9, 2018
    Posted on February 09, 2018 at 16:52

    Please check the note (4) in the datasheet, which says: 'Time to obtain valid data after exiting power-down mode'.

    So you need to provide the turn on time after you set the ODR in CTRL_REG1 register.

    Visitor II
    February 10, 2018
    Posted on February 10, 2018 at 06:34

    Thanks Miroslav. It clears my doubt.